What is a PMMC?

Encyclopedia
Encyclopedia
Field: Encyclopedia
0
China

PMMC Meter Definition


A PMMC meter (also known as a D’Arsonval meter or galvanometer) is defined as a device that measures the current through a coil by observing the coil’s angular deflection in a uniform magnetic field.

 

56d86c511b9534fc13b161aa4646bb3e.jpeg

 

PMMC Construction


A PMMC meter (or D’Arsonval meters) is constructed of 5 main components:


  • Stationary Part or Magnet System

  • Moving Coil

  • Control System

  • Damping System

  • Meter


Working Principle


A PMMC meter uses Faraday’s Laws of electromagnetic induction, where a current-carrying conductor in a magnetic field experiences a force proportional to the current, moving a pointer on a scale.


PMMC Torque Equation


Let us derive a general expression for torque in permanent magnet moving coil instruments or PMMC instruments. We know that in moving coil instruments the deflecting torque is given by the expression:


  • Td = NBldI where N is number of turns,

  • B is magnetic flux density in air gap,

  • l is the length of moving coil,

  • d is the width of the moving coil,

  • I is the electric current.


Now for a moving coil instrument the deflecting torque should be proportional to current, mathematically we can write Td = GI. Thus on comparing we say G = NBIdl. At steady state we have both the controlling and deflecting torques are equal. Tc is controlling torque, on equating controlling torque with deflection torque we have ,GI = K.x where x is deflection thus current is given by

 

de4df743f375d93cf9226fd50a822703.jpeg

 

Since the deflection is directly proportional to the current therefore we need a uniform scale on the meter for measurement of current.

 


Now we are going to discuss about the basic circuit diagram of the ammeter. Let us consider a circuit as shown below:

 

000c792a406fb23fedd52235536ad4ed.jpeg

 

The current I splits into two components at point A: Is and Im. Before discussing their magnitudes, let’s understand shunt resistance construction. The main properties of shunt resistance are detailed below:


The electrical resistance of these shunts should not differ at higher temperature, it they should posses very low value of temperature coefficient. Also the resistance should be time independent. Last and the most important property they should posses is that they should be able to carry high value of current without much rise in temperature. Usually manganin is used for making DC resistance. Thus we can say that the value of Is much greater than the value of Im as resistance of shunt is low. From the we have,

 

56c4f1c985e4ee7328145623c45488ca.jpeg

 

Where, Rs is resistance of shunt and Rm is the electrical resistance of the coil.

98e214baa4027476eaaf675a9ac9df13.jpeg

From the above two equations we can write,

fb51b5ab6175479aa97dcf0851ba4919.jpeg

Where, m is the magnifying power of the shunt.


 

Errors in Permanent Magnet Moving Coil Instruments


  • Errors due to permanent magnets


  • Change in the resistance of the moving coil with the temperature


Advantages of Permanent Magnet Moving Coil Instruments


  • The scale is uniformly divided as the current is directly proportional to deflection of the pointer. Hence it is very easy to measure quantities from these instruments.



  • Power consumption is also very low in these types of instruments.



  • A high torque to weight ratio.



  • These are having multiple advantages, a single instrument can be used for measuring various quantities by using different values of shunts and multipliers.


Disadvantages of Permanent Magnet Moving Coil Instruments


  • These instruments cannot measure AC quantities.

  • The cost of these instruments is high as compared to moving iron instruments.

 

Give a tip and encourage the author!
Recommended
What are the safety precautions and guidelines for using AC load banks?
What are the safety precautions and guidelines for using AC load banks?
AC load banks are electrical devices used to simulate real-world loads and are widely applied in power systems, communication systems, automation control systems, and other fields. To ensure personal and equipment safety during use, the following safety precautions and guidelines must be observed:Select an appropriate AC load bank: Choose an AC load bank that meets actual requirements, ensuring its capacity, voltage rating, and other parameters satisfy the intended application. Additionally, sel
Echo
11/06/2025
What should be noted when installing a Type K thermocouple?
What should be noted when installing a Type K thermocouple?
Installation precautions for Type K thermocouples are critical to ensuring measurement accuracy and extending service life. Below is an introduction to the installation guidelines for Type K thermocouples, compiled from highly authoritative sources:1.Selection and Inspection Select the appropriate thermocouple type: Choose the right thermocouple based on the temperature range, medium properties, and required accuracy of the measurement environment. Type K thermocouples are suitable for temperatu
James
11/06/2025
 Causes and Preventive Measures of Fire and Explosion in Oil Circuit Breakers
Causes and Preventive Measures of Fire and Explosion in Oil Circuit Breakers
Causes of Fire and Explosion in Oil Circuit Breakers When the oil level in an oil circuit breaker is too low, the oil layer covering the contacts becomes too thin. Under the effect of the electric arc, the oil decomposes and releases flammable gases. These gases accumulate in the space beneath the top cover, mixing with air to form an explosive mixture, which can ignite or explode under high temperature. If the oil level inside the tank is too high, the released gases have limited space to expan
Felix Spark
11/06/2025
THD Measurement Error Standards for Power Systems
THD Measurement Error Standards for Power Systems
Error Tolerance of Total Harmonic Distortion (THD): A Comprehensive Analysis Based on Application Scenarios, Equipment Accuracy, and Industry StandardsThe acceptable error range for Total Harmonic Distortion (THD) must be evaluated based on specific application contexts, measurement equipment accuracy, and applicable industry standards. Below is a detailed analysis of key performance indicators in power systems, industrial equipment, and general measurement applications.1. Harmonic Error Standar
Edwiin
11/03/2025
Send inquiry
Download
Get the IEE Business Application
Use the IEE-Business app to find equipment, obtain solutions, connect with experts, and participate in industry collaboration anytime, anywhere—fully supporting the development of your power projects and business.