Why is only the half deflection method used to find the resistance of a galvanometer?

Encyclopedia
Field: Encyclopedia
0
China

Principles of the Half-Deflection Method

The Half-Deflection Method is a technique used to measure the internal resistance of a galvanometer. This method is adopted primarily because it is relatively simple and does not require additional precision instruments or complex circuit setups. Below are the reasons for using the half-deflection method along with its underlying principles:

Basic Principles of the Half-Deflection Method

Initial State: First, connect the galvanometer to a known voltage source so that the galvanometer needle reaches full-scale deflection. At this point, the current through the galvanometer is I, assuming the internal resistance of the galvanometer is G.

Adding Resistance: Next, place a variable resistor R in series with the galvanometer branch and adjust this resistor until the galvanometer needle deflects to half its original position. At this moment, the current through the galvanometer has decreased to I/2.

Calculating Galvanometer Resistance: According to Ohm's Law and the voltage divider principle, when the galvanometer's deflection is halved, the voltage across the galvanometer also halves. Therefore, it can be concluded that the voltage across the galvanometer VG equals the voltage across the external resistor VR. Since the current has also been halved, we have:

image.png

This means that when the galvanometer's deflection is halved, the value of the external resistor R is equal to the internal resistance G of the galvanometer.

Why Only Use the Half-Deflection Method?

  • Simplified Measurement Process: The half-deflection method requires only a simple experimental setup—a power supply, a galvanometer, and a variable resistor. It does not require precise knowledge of the power supply voltage or the exact current values; it only needs to observe changes in the galvanometer needle.

  • Avoiding Complex Calculations: Compared to other methods, such as the Wheatstone Bridge method, the half-deflection method does not involve complex balance conditions or equation solving, making it easier to understand and operate.

  • Direct Reading: By directly observing the position changes of the galvanometer needle, results can be obtained quickly, reducing measurement time and sources of error.

  • Suitable for Educational Demonstrations: For students, this is an intuitive and easy-to-understand method that helps grasp basic circuit concepts and physical laws, such as Ohm's Law and the voltage divider rule.

However, it is important to note that while the half-deflection method is convenient and quick, it also has limitations. For example, if the galvanometer itself exhibits nonlinear response characteristics, this method may not be accurate enough. Additionally, this method relies on visual judgment of the galvanometer needle, which might introduce human error. Therefore, in cases where higher precision is required, more accurate methods such as the aforementioned Wheatstone Bridge method or other high-precision testing equipment are typically used.

Give a tip and encourage the author!

Recommended

Classification of Equipment Defects for Relay Protection and Safety Automatic Devices in Substations
In daily operations, various equipment defects are inevitably encountered. Whether maintenance personnel, operation and maintenance staff, or specialized management personnel, all must understand the defect classification system and adopt appropriate measures according to different situations.According to Q/GDW 11024-2013 "Operation and Management Guide for Relay Protection and Safety Automatic Devices in Smart Substations," equipment defects are classified into three levels based on severity an
12/15/2025
Under What Conditions Will the Line Circuit Breaker Auto-Reclosing Signal Be Locked Out?
The line circuit breaker auto-reclosing signal will be locked out if any of the following conditions occur:(1) Low SF6 gas pressure in circuit breaker chamber at 0.5MPa(2) Insufficient energy storage in circuit breaker operating mechanism or low oil pressure at 30MPa(3) Busbar protection operation(4) Circuit breaker failure protection operation(5) Line distance protection zone II or zone III operation(6) Short lead protection operation of circuit breaker(7) Presence of remote tripping signal(8)
12/15/2025
Application of Auto-Reclosing Residual Current Protective Devices in Lightning Protection for Communication Power Supplies
1. Power Interruption Problems Caused by RCD False Tripping During Lightning StrikesA typical communication power supply circuit is shown in Figure 1. A residual current device (RCD) is installed at the power supply input terminal. The RCD primarily provides protection against electrical equipment leakage currents to ensure personal safety, while surge protective devices (SPDs) are installed on power supply branches to protect against lightning intrusions. When lightning strikes occur, the senso
12/15/2025
Reclosing Charge Time: Why Does Reclosing Require Charging? What Effects Does Charging Time Have?
1. Function and Significance of Reclosing ChargingReclosing is a protective measure in power systems. After faults such as short circuits or circuit overloads occur, the system isolates the faulty circuit and then restores normal operation through reclosing. The function of reclosing is to ensure continuous operation of the power system, improving its reliability and safety.Before performing reclosing, the circuit breaker must be charged. For high-voltage circuit breakers, the charging time is g
12/15/2025
Send inquiry
+86
Click to upload file
Download
Get the IEE Business Application
Use the IEE-Business app to find equipment, obtain solutions, connect with experts, and participate in industry collaboration anytime, anywhere—fully supporting the development of your power projects and business.